Health & Medical Hematopathy & blood disease

How to Increase Pressure in a Flushometer

    • 1). Remove the vandal-resistant cap from the Bak-Chek valve sleeve. Insert the tip of a slot screwdriver under the edge of the cap and work it around the cap, prying the cap away from the Bak-Chek valve sleeve until it pops off.

    • 2). Rotate the control-stop screw underneath the cap counterclockwise with a screwdriver to open the control valve completely. Test the flushing action to see if it is sufficient.

    • 3). Turn the control valve all the way clockwise to close the valve completely if the flush volume is still inadequate.

    • 4). Remove the large diaphragm cover on the top of the unit with a pipe wrench or slip-joint pliers.

    • 5). Remove the relief valve that is installed in the center of the diaphragm by lifting it completely out.

    • 6). Reinstall the diaphragm cover, open the control valve all the way and test the flushing action to see if it is sufficient.

    • 7). Turn the control valve clockwise all the way to close the valve completely if the flush volume is still inadequate.

    • 8). Remove the large diaphragm cover on the top of the unit with a pipe wrench or slip-joint pliers.

    • 9). Lift the diaphragm assembly out of the diaphragm bowl as a single unit.

    • 10

      Flip the refill head that is underneath the diaphragm in the diaphragm bowl to convert the Flushometer from a 3.5 gallon-per-flush unit to a 4.5 gallon-per-flush unit.

    • 11

      Reinstall the diaphragm assembly into the diaphragm bowl. Reinstall the diaphragm cover, open the control valve all the way and test the flushing action.

SHARE
RELATED POSTS on "Health & Medical"

Leave Your Reply

*